Precautions for Calibration of the VN3100 Portable Turbidimeter
VVNA Solutions:
Calibration of the VN3100 Portable Turbidimeter is of great importance, as it directly determines the accuracy of measurement results. The precautions for calibration are as follows:
Shake the required standard solutions thoroughly (for 2–3 minutes).
Let the well-shaken standard solutions stand for 5 minutes.
Thoroughly clean, rinse and dry the outer walls of the reagent bottles.
Before using all standard solutions, quickly apply a layer of silicone oil to the outer walls of the vials.
Perform the calibration procedure.