Precautions for Calibration of the VN3100 Portable Turbidimeter

Precautions for Calibration of the VN3100 Portable Turbidimeter

Precautions for Calibration of the VN3100 Portable Turbidimeter
VVNA Solutions:

Calibration of the VN3100 Portable Turbidimeter is of great importance, as it directly determines the accuracy of measurement results. The precautions for calibration are as follows:

  1. Shake the required standard solutions thoroughly (for 2–3 minutes).

  2. Let the well-shaken standard solutions stand for 5 minutes.

  3. Thoroughly clean, rinse and dry the outer walls of the reagent bottles.

  4. Before using all standard solutions, quickly apply a layer of silicone oil to the outer walls of the vials.

  5. Perform the calibration procedure.