Water Quality Encyclopedia | Unstable COD Test Results? How Many of These 8 Hidden Pitfalls Have You Fallen Into?
-
2025-12-17
-
Click Count:763
Editor:VVNA
Source:Shenzhen TongAo Technology Co., Ltd
When performing rapid COD (Chemical Oxygen Demand) tests, have you often encountered this dilemma: you follow the standard procedure step by step, yet the test results for the same water sample fluctuate drastically, varying by tens or even hundreds of mg/L? You recalibrate the instrument repeatedly, prepare new reagents, and waste hours troubleshooting, only to find no root cause. This not only wastes time and consumables but may also lead to misleading water quality assessments.
In fact, 90% of COD test result fluctuations are not caused by errors in core steps, but by overlooked "invisible pitfalls". This guide will break down the 8 most common mistakes, covering everything from equipment preparation to final reading, helping you stabilize your COD data comprehensively!
Pitfall 1: Using "Any Available Tube" Instead of Specialized Digestion Colorimetric Tubes
Many people opt for convenience by using ordinary sealed glass tubes instead of specialized digestion colorimetric tubes, thinking "they are just containers for solutions". However, this practice hides significant risks. Ordinary glass tubes have inconsistent wall thickness and light transmittance, which directly affect the accuracy of colorimetric readings. Worse still, the glass may adsorb organic substances, leading to distorted results. Additionally, their insufficient heat resistance can cause deformation or even rupture during high-temperature and high-pressure digestion at 165℃, posing safety hazards and potentially causing liquid leakage and volatilization.
Correct Practice: You must use 16mm specialized sealed digestion colorimetric tubes that are compatible with your instrument. Always use the dedicated tubes provided by the manufacturer. Their material, sealing performance, and optical properties are all custom-designed for optimal compatibility, serving as the first line of defense for accurate data.
Pitfall 2: Forgetting to Shake the Water Sample Before Sampling
According to the specifications in HJ/T399-2007 Water Quality - Determination of Chemical Oxygen Demand - Rapid Digestion Spectrophotometric Method and HJ828—2017 Water Quality - Determination of Chemical Oxygen Demand - Dichromate Method (which replaces GB11914-89), COD tests measure the total oxygen consumption of both soluble and suspended organic matter in water samples. If the water sample is left undisturbed for too long before sampling, suspended solids will settle to the bottom, and you will only collect the "upper clear liquid". This will result in a systematically low test value, with deviations easily exceeding 10% — rendering even the most precise subsequent operations useless.
Suspended solids encapsulate organic matter. In some water samples, the suspended particles vary significantly in size, and there may even be visible impurities. Even after shaking, when using a pipetting tool to measure the liquid, the suspended solids in the collected sample may still be unevenly distributed, leading to large result deviations.
Correct Practice: Before sampling, hold the water sample bottle with both hands and invert it up and down 10-15 times. For water samples with a large amount of sediment, you can extend the shaking time or use a stirrer for assistance. Ensure that the suspended solids are uniformly distributed, making the sample representative.
Pitfall 3: Determining Cooling Time "By Feel"
Rushing to test the sample while it is still hot after digestion will cause errors, as high temperatures alter the solution density, leading to changes in the colorimetric optical path and inaccurate absorbance readings. On the other hand, allowing the sample to cool completely naturally can take 1-2 hours, especially in high room temperatures, seriously slowing down efficiency.
Correct Practice: The digestion tube must be cooled to room temperature (approximately 25℃) before testing. You can first cool it in a fume hood for 15 minutes, then place it in a constant temperature water bath at 25℃ for 5 minutes. This method balances accuracy and efficiency. The above times can be adjusted according to the on-site environment.
Pitfall 4: Shaking the Colorimetric Tube "Out of Habit" Before Reading
After cooling, if you see sediment at the bottom of the tube, it is a common habit to shake the colorimetric tube. However, you should be aware that components such as chlorine interference inhibitors in specialized COD reagents are insoluble substances. Shaking will resuspend them, interfering with the optical path and causing fluctuations in absorbance, which results in extremely large data deviations.
Correct Practice: Before loading the tube into the instrument for testing, always hold it up to the light for inspection. The solution must be colored but clear and transparent. If turbidity appears or the solution becomes turbid after shaking, you can place the tube back into the digester and use the residual heat to quickly settle the suspended solids within 2-3 minutes. As long as the sediment does not cover the optical path area, it will not affect the results.
Pitfall 5: Treating the Blank Experiment "Perfunctorily"
Many people view the blank experiment as merely a "formality". They casually use tap water, mineral water, or uncalibrated distilled water for the blank, or even skip this step entirely. However, they overlook that the blank is crucial for eliminating reagent impurities and instrument baseline errors. If the blank water sample contains trace amounts of organic matter or interfering substances, it will lead to a high blank value, resulting in systematic distortion of the final calculated COD value.
Correct Practice: You must prepare the blank sample using ultrapure water from the same batch as the experimental water (commercially available purified water such as Wahaha or C'estbon can be used as alternatives). Additionally, the reagent dosage and digestion conditions for the blank tube must be identical to those for the sample tubes. During testing, always calibrate with the blank first, then measure the samples to ensure that errors are offset.
Pitfall 6: Diluting the Water Sample "By Empirical Estimation"
High-concentration water samples need to be diluted before testing. Many people skip the preliminary test and arbitrarily dilute the sample based on experience. This often results in two problems: either the dilution factor is insufficient, causing the result to exceed the instrument's measurement range and leading to saturated and distorted readings; or the dilution factor is too high, amplifying the introduced errors and causing the result to deviate significantly from the true value.
Correct Practice: First, perform a preliminary test to determine the approximate concentration range. Then, select the appropriate dilution factor based on the principle of "ensuring the test result falls within the middle section of the instrument's measurement range". When diluting, use a calibrated pipette to ensure the accuracy of the dilution ratio.
Pitfall 7: Cleaning Equipment "Superficially"
Inadequate cleaning of equipment such as digestion tubes and pipettes can lead to contamination of new samples by residual organic matter or reagents from previous tests. In particular, if residual chlorine from chlorine-containing disinfectants is not rinsed off completely, it will react with the COD reagents, causing the test results to be abnormally high.
Correct Practice: First, rinse the equipment with tap water. Then, soak it in dilute nitric acid for 10 minutes. Finally, rinse it with distilled water at least three times and allow it to dry before use. Avoid cleaning with chlorine-containing disinfectants or hydrochloric acid. If they must be used, additional rinsing steps are required.
Pitfall 8: Not Waiting Long Enough for Instrument Warm-Up
Conducting tests immediately after turning on the instrument will lead to unstable readings, as the instrument's optical system and circuit have not yet stabilized. This causes drift in absorbance readings and poor consistency in test results for the same batch of samples.
Correct Practice: After turning on the instrument, it needs to warm up for 5-10 minutes (warm-up time varies by manufacturer; some instruments require no warm-up, while others require 30 minutes or more). During this period, you can perform blank calibration. Start testing only when the instrument displays stable readings. Avoid frequently turning the instrument on and off or moving it during the testing process.
Summary: 8 Key Details to Stabilize COD Test Results
The stability of COD test results lies entirely in the operational details:
Equipment Compatibility: Use specialized digestion colorimetric tubes — never settle for less;
Sample Homogenization: Shake the sample thoroughly before sampling to ensure representativeness;
Adequate Cooling: Cool the sample to room temperature before testing — do not rush;
Static Reading: Do not shake the colorimetric tube before reading to avoid optical path interference;
Standardized Blank: Use ultrapure water for the blank sample and strictly synchronize operations;
Accurate Dilution: Perform a preliminary test first to determine the optimal dilution factor;
Thoroughly Clean Equipment: Ensure no residues remain to avoid cross-contamination;
Instrument Warm-Up: Allow the instrument to stabilize for 10 minutes after startup before testing.
What other "unexpected pitfalls" have you encountered in COD testing? Were they small mistakes in reagent preparation or minor incidents in instrument operation? Feel free to leave us a message. Let's avoid these pitfalls together and make water quality testing data more stable and accurate!
Scan the QR code with WeChat to share with friends
This article is reprinted from cooperative media, other websites on the Internet, or partially edited and created by VVNA. VVNA publishes this article for the purpose of disseminating more information, which does not mean endorsing its views or verifying its descriptions. If you have any questions about the content, please contact VVNA customer service.
All content reprinted on this website comes from reliable sources and is clearly marked with its origin. (If the information of the original author is unavailable when reprinting, the author's name may not be indicated.) The copyright of the content shall belong to the original author. Due to objective constraints, if we fail to get in touch with the author, or the author disagrees with the publication of the content on this website, or any errors are found, please contact us. We will respect the author's wishes and make corrections or deletions in a timely manner.
Please be sure to respect the copyright and authorship of the works when reprinting or downloading them.
The content of the article is for reference only. Please indicate the source if reprinted.
-
- Follow VVNA on WeChat
Stay updated on VVNA's updates
-
Emergency Water Quality Testing Solutions -
Aquaculture Water Quality Testing Solutions -
Pool Water Quality Testing Solutions -
Municipal Sewage Water Quality Testing Solutions -
Industrial Water Quality Testing Solutions -
Application of the VN3100 Portable Turbidimeter in Winemaking Facilities Swimming Pool and SPA Water -
Application of the VN3100 Portable Turbidimeter in Municipal Wastewater Treatment -
Application of the VN3100 Portable Turbidimeter in the Hydrology and Water Resources Sector

