How can the error caused by the sample cuvette be reduced for the VN3100 Portable Turbidimeter?
-
Update:2025-07-21
-
Popularity:1912
Answer Source:VVNA
How can the error caused by the sample cuvette be reduced for the VN3100 Portable Turbidimeter?
VVNA Solutions:
How to Reduce Errors Caused by the Sample Cell in the VN3100 Portable Turbidimeter?
When the Sample Cell Surface Is Contaminated:
Keep the sample cell clean by washing it with detergent, acid, or other suitable cleaning agents;
Perform ultrasonic cleaning;
Rinse the sample cell surface with high-purity water.
Ensure the sample cell remains nearly transparent;
When the Sample Cell Has Scratches:
To eliminate or reduce the impact of scratches on measurement results:
Wipe the sample cell surface with silicone oil;
Mark the sample cell and use the same position for each measurement.