How can the error caused by air bubbles be eliminated for the VN3100 Portable Turbidimeter?

  • Update:2025-07-21
  • Popularity:1906
  • Answer Source:VVNA

How can the error caused by air bubbles be eliminated for the VN3100 Portable Turbidimeter?

How can the error caused by air bubbles be eliminated for the VN3100 Portable Turbidimeter?

How can the error caused by air bubbles be eliminated for the VN3100 Portable Turbidimeter?
VVNA Solutions:

The presence of air bubbles scatters light, leading to higher reading values.

Sample Degassing Methods:

  1. Let the sample stand undisturbed for a period of time to eliminate air bubbles.

  2. Perform ultrasonic treatment.

  3. Apply vacuum extraction.